CLARION Area Control Panel
Sliding glass front · flush, wall-aligned design
The unit consolidates medical gas lines from wards, corridors, and operating areas into a single control panel, providing both monitoring and isolation functions. Each line is equipped with a dedicated manometer, low-pressure sensor, and medical gas shut-off valve. It can be configured with or without alarms according to project requirements. Its recessed installation and smooth sliding glass panel design create a minimal, modern, and sophisticated appearance within the space. The glass surface can be customized with colors, patterns, or graphics to achieve a project-specific aesthetic. Supports 1–6 gas lines.
Supports 1–6 gas lines; panel width increases with the number of lines. Each line includes manometer and pressure sensor for continuous monitoring. Data can be sent to the Building Management System (BMS) for central monitoring.
KEY FEATURES
- Gas Lines — 1–6 Gas Lines, Project-Specific Setup
- Custom Surface Application — Custom design on the front panel.
- Instant Emergency Supply — Maintains continuous gas flow
- Seamless Surface Effect — Wall-aligned, continuous appearance.
- Controlled Sliding Mechanism — Smooth front panel movement.
- Alarm Options — Touch Panel / Digital / No Alarm
TECHNICAL SPECIFICATIONS
- Supply Voltage
- 220 V, 50 Hz.
- Operating Temperature
- 0°C – 40°C
- Relative Humidity
- 10 – 95 %
- Number of Lines (Max.)
- 1 – 6 Lines
- Medical Gases
- O₂ – N₂O – MA4 – SA7 – VAC – CO₂ – N₂
- Configuration
- Number of lines and gas combination are project-specific
- Max. Inlet Pressure
- 10 Bar
- Operating Pressure
- 0 – 8.5 Bar (Compressed Gas) 0–(-1) Bar (Vacuum)
- Pressure Sensitivity
- 0.1 Bar
- Inlet/Outlet Pipes
- Top Entry, Medical Copper Pipe
- Pipe Diameter
- Ø15 mm (Pressure Lines) Ø22 mm (Vacuum Line)
- Pressure Gauges
- 0–10 Bar
- Vacuum Gauge
- 0–(-1) Bar
- Emergency Supply
- Gas-specific NIST connections (Project-specific gas outlet option)
- Line Control
- Ball Valve
- Monitoring
- Pressure Gauge –
- Material
- Cold Rolled Steel (Lower Body), Glass Front Panel
- Installation
- Flush Mounted / Surface Mounted Recessed
- Recommended Installation Height
- 1550 – 1650 mm from Floor Level (Panel Center)
- Dimensions (H×W×D)
- 128 × 178 × 68 mm
- Control
- Microprocessor Controlled Electronic System
- Digital Alarm Display
- 7-Segment Digital Display
- Touchscreen Alarm Display
- 5" Touchscreen Display
- Sensor Type
- Pressure and Vacuum
- Alarm Function
- High/Low Pressure Alarm, Empty Gas Cylinder, Intermediate Pressure Alarm, System Changeover Alarm
- Alarm Notification
- Audible and Visual Warning + Notification to Master Alarm Panel
- Management Standards
- 2007/42/EC
- Quality Management System
- EN ISO 13485:2016 — EN ISO 9001:2015
- Electrical Safety and Performance
- EN 60601-1:2009 — EN 60601-1-2:2011 — EN 60601-1-8:2008
- Risk, Usability and Labeling
- EN ISO 14971 :2012 — IEC62366-1:2015 EN ISO 15223-1:2016 — EN 1041+A1:2011
Technical specifications are subject to change without notice. For technical drawings, model codes and all configurations, see the product catalogue (PDF) .